The publication detail shows the title, authors (with indicators showing other profiled authors), information on the publishing organization, abstract and a link to the article in PubMed. This abstract is what is used to create the fingerprint of the publication. If any grants are referenced by the publication, they will be listed here as well.



Anatomic stereotactic catheter ablation on three-dimensional magnetic resonance images in real time.

Timm Dickfeld; Hugh Calkins; Muz Zviman; Ritsushi Kato; Glenn Meininger; Lars Lickfett; Ron Berger; Henry Halperin; Stephen B Solomon (Profiled Authors: Ronald Berger; Timm-Michael Dickfeld; Hugh Calkins; Henry Halperin; Menekhem Zviman)

Division of Cardiology, Johns Hopkins School of Medicine, Baltimore, Md, USA. tdickfel@jhmi.edu
Circulation 2003;108(19):2407-13.

Abstract

BACKGROUND: Targets for radiofrequency (RF) ablation of atrial fibrillation, atrial flutter, and nonidiopathic ventricular tachycardia are increasingly being selected on the basis of anatomic considerations. Because fluoroscopy provides only limited information about the relationship between catheter positions and cardiac structures and is associated with radiation risk, other approaches to mapping may be beneficial. METHODS AND RESULTS: An electromagnetic catheter positioning system was superimposed on 3D MR images using fiducial markers. This allowed the dynamic display of the catheter position on the true anatomy of previously acquired MR images in real time. In vitro accuracy and precision during catheter navigation were assessed in a phantom model and were 1.11+/-0.06 and 0.30+/-0.07 mm (mean+/-SEM), respectively. Left and right heart catheterization was performed in 7 swine without the use of fluoroscopy, yielding an in vivo accuracy and precision of 2.74+/-0.52 and 1.97+/-0.44 mm, respectively. To assess the reproducibility of RF ablation, RF lesions were created repeatedly at the identical anatomic site in the right atrium (n=8 swine). Average distance of the repeated right atrial ablations was 3.92+/-0.5 mm. Straight 3-point lines were created in the right and left ventricles to determine the ability to facilitate complex ablation procedures (n=6 swine). The ventricular lesions deviated 1.70+/-0.24 mm from a straight line, and the point distance differed by 2.25+/-0.63 mm from the pathological specimen. CONCLUSIONS: Real-time display of the catheter position on 3D MRI allows accurate and precise RF ablation guided by the true anatomy. This may facilitate anatomically based ablation procedures in, for instance, atrial fibrillation or nonidiopathic ventricular tachycardia and decrease radiation times.

Scientific Context

This section shows information related to the publication - computed using the fingerprint of the publication - including related publications, related experts and related grants with fingerprints representing significant amounts of overlap between their fingerprint and this publication. The red dots indicate whether those experts or terms appear within the publication, thereby showing potential and actual connections.

Related Publications