A comparison of radiographic techniques and electromagnetic transponders for localization of the prostate

Ryan D. Foster; David A. Pistenmaa; Timothy D. Solberg

(Profiled Authors: Ryan D Foster; David A Pistenmaa; Timothy D Solberg)

Radiation Oncology. 2012;7(1):[101].

Abstract

Background: The aim of this study is to compare three methodologies of prostate localization and to determine if there are significant differences in the techniques.Methods: Daily prostate localization using cone beam CT or orthogonal kV imaging has been performed at UT Southwestern Medical Center since 2006. Prostate patients are implanted with gold seeds, which are matched with the planning CT or DRR before treatment. More recently, a technology using electromagnetic transponders implanted within the prostate was introduced into our clinic (Calypso®). With each technology, patients are localized initially using skin marks and the room lasers. In this study, patients were localized with Calypso and either CBCT or kV orthogonal images in the same treatment session, allowing a direct comparison of the technologies. Localization difference distributions were determined from the difference in the offsets determined by CBCT/kV imaging and Calypso. CBCT-Calypso and kV imaging-Calypso localization data were summarized from over 900 and 250 fractions each, respectively. The Wilcoxon signed rank test is used to determine if the localization differences are statistically significant. We also calculated Pearson's product-moment correlation coefficient (R2) to determine if there is a linear relationship between the shifts determined by Calypso and the radiographic techniques.Results: The differences between CBCT-Calypso and kV imaging-Calypso localizations are -0.18 ± 2.90 mm, -0.79 ± 2.18 mm, -0.01 ± 1.20 mm and -0.09 ± 1.40 mm, 0.48 ± 1.50 mm, 0.08 ± 1.04 mm, respectively, in the AP, SI, and RL directions. The Pearson product-moment correlation coefficients for the CBCT-Calypso shifts were 0.71, 0.92 and 0.88 and for the OBI-Calypso comparison were 0.95, 0.89 and 0.85. The percentage of localization differences that were less than 3 mm were 86.1%, 84.5% and 96.0% for the CBCT-Calypso comparison and 95.8%, 94.3% and 97% for the kV OBI-Calypso comparison. No trends were observed in the Bland-Altman analysis.Conclusions: Localization of the prostate using electromagnetic transponders agrees well with radiographic techniques and each technology is suitable for high precision radiotherapy. This study finds that there is more uncertainty in CBCT localization of the prostate than in 2D orthogonal imaging, but the difference is not clinically significant. © 2012 Foster et al.; licensee BioMed Central Ltd.


PMID: 22720845     PMCID: PMC3431985    

Scientific Context

This section shows information related to the publication - computed using the fingerprint of the publication - including related publications, related experts with fingerprints representing significant amounts of overlap between their fingerprint and this publication. The red dots indicate whether those experts or terms appear within the publication, thereby showing potential and actual connections.

Related Publications

Related Topics

Appears in this Publication Appears in this Document

Related Experts

Author of this Publication Author of this Document